Application
Location:Home - Application - Detection of Phthalate Compounds in Food
Time:2025-12-25

Instruments and equipment
Gas chromatography-mass spectrometry (GC1290/MS8100, Sunny Hengping Instrument), equipped with capillary split/non split injection ports (full EPC), constant current and constant pressure functions, column temperature box with 20 stage programmed heating, and EI source.
Column 2: DB-5MS (30m * 0.25mm * 0.25um)
3 162 position reverse control automatic sampler (Swiss CTC)
4 Electronic Analytical Balance (AE224, Sunny Hengping Instrument)
5 helium gas (purity 99.999%)
Sixteen mixed standard samples of phthalates, including dimethyl phthalate (DMP), diethyl phthalate (DEP), and diisobutyl phthalate (DIBP)
)Dibutyl phthalate (DBP), di (2-methoxy) ethyl phthalate (DMEP), di (4-methyl-2-pentyl) phthalate (BMPP), di (2-ethoxy) ethyl phthalate (DEEP), di pentyl phthalate (DPP), di hexyl phthalate (DHXP), butyl benzyl phthalate (BBP), di (2-butoxyethyl) phthalate (DBEP), di (2-ethylhexyl) phthalate (DEHP), di (2-ethylhexyl) phthalate Esters (DPHP), di-n-octyl phthalate (DNOP), and dinonyl phthalate (DNP).
Experimental conditions
1. Injection port temperature: 250 ℃
2. Program heating: Initial temperature of 60 ℃, hold for 1 minute; Heat up to 220 ℃ at a rate of 20 ℃/min and maintain for 1 minute; Heat up to 280 ℃ at a rate of 5 ℃/min and maintain for 3 minutes.
3 Injection methods: Non split injection
4 Injection volume: 1μ l
5 ion source temperature: 230 ℃
6 Transmission line temperature: 280 ℃
7 Scanning Method: Select ion scanning and monitor ions according to Appendix B of GB 5009.271-2016
8 solvent delay: 7 minutes